Previous revision
commands:aflim [2018/02/07 17:15] – joncommands:aflim [2026/08/31 17:36] (current) – tag order brandon
Line 1: Line 1:
 +====== Command:AFLIM (AL) ======
  
 +--> For CRISP ^
 +
 +--> MS2000 and RM2000 Syntax ^
 +
 +^Shortcut |AL|
 +^Format  |AFLIM [X=Log_amp_AGC] [Y=LED_intesity_pot] [Z=in_focus_mm]   |
 +^Remembered |Using SS Z |
 +
 +<--
 +
 +--> Tiger Syntax ^
 +
 +^Shortcut |AL|
 +^Format  |[Addr#]AFLIM [X=Log_amp_AGC] [Y=LED_intesity_pot] [Z=in_focus_mm]   |
 +^Type |Card-Addressed|
 +^Remembered |Using [Addr#]SS Z |
 +
 +<--
 +
 +**X/Y:** The X and Y arguments of this command to directly read and write values (0 to 255) to the CRISP electronics digital potentiometers.  <wrap info>Not recommended for use with host software.</wrap>
 +
 +  * See the [[:commands:lock|LK M]] command to adjust the LogAmp_AGC instead of ''AL X''.
 +  * See the [[:commands:unlock|UL X]] command to adjust the LED intensity instead of ''AL Y''.
 +\\
 +**Z** ''[in_focus_mm]'': The Z-argument specifies the focus precision (in millimeters) when the lock state changes from K or k to F. Useful for automatic checking of desired focus stability.  Also useful to enforce a tighter or looser focus state before indicating a lock condition.  Note that this value is overwritten whenever the NA of the objective is specified via the LR Y command as of November 2015.
 +
 +<--
 +
 +--> For Video Autofocus ^
 +
 +--> MS2000 and RM2000 Syntax ^
 +
 +^Shortcut |AL|
 +^Format  |AFLIM [X= x-axis highlight] [Y= y-axis highlight] [Z= safety limit enable]  |
 +^Remembered |Using SS Z |
 +
 +<--
 +
 +--> Tiger Syntax ^
 +
 +^Shortcut |AL|
 +^Format  |[Addr#]AFLIM [X= x-axis highlight] [Y= y-axis highlight] [Z= safety limit enable]  |
 +^Type |Card-Addressed|
 +^Remembered |Using [Addr#]SS Z |
 +
 +<--
 +
 +**X/Y:** The X and Y values set the length and breadth of the Sampled/Highlighted Video area.  Range is 0 to 100, with the value of 0 covering 0% of the video frame and 100 covering 90% of video frame. 
 +
 +**Z:** The Z value enables or disables the 200 µm safety limit described in the AUTOFOCUS OPERATION section on page 4.  Setting safety limit enable = 1 enables the safety limit; safety limit enable = 0 disables the safety feature.  The default value is 1.  
 +
 +<WRAP center round important 60%>
 +**Warning!** Disabling the safety limit could result in damage to your optics, your sample, or your focus drive.
 +</WRAP>
 +
 +<asi>$ AL X=80 Y=50 Z=1
 +:A<CR><LF></asi>
 +<asi>$ AL
 +:N-3 </asi>
 +Error indicates missing arguments
 +<asi>$ AL X=1000 Y=-12
 +:N-4</asi>
 +Error indicates arguments out of range
 +<asi>$ AL X=90 Y=90
 +:N-5 </asi>
 +Error indicates operation failed, try entering one argument at a time
 +<asi>$ AL X? Y? Z?
 +:A X=80 Y=50 Z=1</asi>
 +
 +
 +<--
 +
 +
 +
 +
 +{{tag>crisp autofocus commands tiger ms2000}}