| + | ====== Leitz LaborLux D Microscope XY Stage Installation ====== | ||
| + | This procedure steps you through the installation and alignment of the MC-XYZ xy stage onto the Laborlux-D microscope. The following tools are required for this procedure: | ||
| + | |||
| + | small slotted screwdriver | ||
| + | 1.5mm hex wrench (provided) | ||
| + | 3mm hex wrench (provided) | ||
| + | |||
| + | ===== XY Stage Installation ===== | ||
| + | |||
| + | The xy stage attaches to the microscope via an adapter ring that is mounted in place of the original xy stage. Installing the ASI adapter ring and stage involves steps: | ||
| + | |||
| + | |||
| + | ==== Step 1 - Removing the condenser and condenser carrier ==== | ||
| + | |||
| + | |||
| + | To remove the existing xy stage the condenser and condenser carrier must first be removed. Remove the condenser by loosening the silver condenser lock screw located on the lower left of the carrier assembly and pull the condenser straight out. (Please refer to the microscope manual if necessary). | ||
| + | |||
| + | |||
| + | |||
| + | ==== Step 2 - Removing the original xy stage ==== | ||
| + | |||
| + | |||
| + | |||
| + | The original xy stage is secured to the microscope with three 3mm allen head screws. | ||
| + | |||
| + | |||
| + | |||
| + | ==== Step 3 - Installing the ASI adapter ring assembly. ==== | ||
| + | |||
| + | |||
| + | |||
| + | Install the ASI adapter ring onto the microscope using the same three mounting holes that the original xy stage used. However, be sure to use the shorter 4x8mm allen head screws that came with the adapter ring to secure it to the microscope. | ||
| + | |||
| + | |||
| + | ==== Step 4 - Installing the xy stage ==== | ||
| + | |||
| + | |||
| + | |||
| + | The xy stage slides over the adapter ring and is secured to the ring with the thumb screw located on the front of the stage. | ||
| + | |||
| + | |||
| + | |||
| + | ==== Step 6 - Reinstalling the condenser carrier and condenser ==== | ||
| + | |||
| + | |||
| + | |||
| + | The installation of the condenser carrier and condenser is in reverse order of step #1. | ||
| + | |||
| + | |||
| + | |||
| + | |||
| + | **Caution please make sure that the electronic/ | ||
| + | ** | ||
| + | |||
| + | {{tag> | ||