ls50
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| ls50 [2016/09/07 18:29] – ↷ Links adapted because of a move operation asiadmin | ls50 [2024/11/26 18:26] (current) – [Introduction] make warning bold brandon | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| + | ====== LS-50 and other Linear Stages ====== | ||
| + | This manual pertains to ASI’s LS-Series linear positioning stages used with the MS2000 or Tiger Multi-card controller systems. | ||
| + | |||
| + | This manual will describe the installation, | ||
| + | |||
| + | |||
| + | ===== Introduction ===== | ||
| + | |||
| + | {{ ls50_1_.png? | ||
| + | |||
| + | {{ ls50_2_.jpg? | ||
| + | |||
| + | * Standard units come with built in rotary encoders | ||
| + | * Linear encoder units are also available for unsurpassed | ||
| + | * Precision ground cross roller bearings and guides are used to | ||
| + | |||
| + | [{{ ls50_1_.jpg? | ||
| + | |||
| + | The LS linear stage provides sub-micron accuracy, and derives precise control through the use of closed-loop DC servomotors employing high-resolution rotary encoders for positioning feedback. An optional linear encoder can also be added to the unit to provide even greater positioning accuracy. The stage utilizes crossed-roller slides, precision lead-screws, | ||
| + | |||
| + | The unit can be configured with a number of lead screw options as outlined in the table below. In our standard rotary encoder configuration, | ||
| + | |||
| + | The optional linear encoder provides a resolution of 20 nm and can be used to provide the ultimate feed back device to increase these factors to a greater degree. Actual resolution & repeatability factors vary somewhat with the load so please contact ASI for these specifications. | ||
| + | |||
| + | The MS-2000 controller accepts industry standard commands, provides automatic backlash correction, and allows RS-232 or USB communication with a host computer. | ||
| + | |||
| + | ^Lead Screw Pitch Options ^Rotary Encoder Resolution ^Maximum Speed <wrap lo> | ||
| + | |6.35 mm |88 nm |7 mm/sec| | ||
| + | |1.57 mm |22 nm |1.75 mm/sec| | ||
| + | |0.317 mm |4.5 nm |0.35 mm/sec| | ||
| + | |||
| + | <WRAP center round important 60%> | ||
| + | **WARNING: | ||
| + | </ | ||
| + | |||
| + | |||
| + | |||
| + | |||
| + | ===== Features and Capabilities of an L-Series with a Controller ===== | ||
| + | |||
| + | |||
| + | *Closed-loop DC servo motor control of the X, Y, Z and Theta axes for precise positioning and highly repeatable focusing | ||
| + | *Sub-micron repeatability on all axes | ||
| + | *Wide dynamic speed range with adjustable trapezoidal move profiles | ||
| + | *Compact ergonomic tabletop control unit is 3½ x 9 x 6½ inches (9 x 23 x 16½ cm) | ||
| + | *Smooth adjustable dual-range joystick control | ||
| + | *Microprocessor control USB communications | ||
| + | *Upper and Lower Maximum Travel Hall-effect limit sensors | ||
| + | *Electronic torque limit on drives minimizes damage by runaway stage | ||
| + | *Other functions including programmable positioning patterns and scans | ||
| + | |||
| + | |||
| + | ===== LS-Series Linear Stages ===== | ||
| + | |||
| + | The LS-50 and LS-100 linear stages provide a simple building block approach to multi-axis motion control when used with ASI’s MS2000 or Tiger controller. | ||
| + | |||
| + | {{ ls50_stage_2_.jpg? | ||
| + | {{ ls50_stage_3_.jpg? | ||
| + | | ||
| + | The photos above show a pair of LS stages set up for XY motion, and an ASI stage mounted to an LS-50 Z-axis unit. The LS-50 could also be mounted directly below the ASI stage as shown below, where the LS-50 is extended to its maximum travel. | ||
| + | | ||
| + | {{ ls50_stage_4_.jpg? | ||
| + | {{ ls50_stage_5_.jpg? | ||
| + | {{ ls50_stage_6_.jpg? | ||
| + | |||
| + | |||
| + | The LS-Series linear stages provide sub-micron accuracy, and derives precise control through the use of closed-loop DC servomotors employing high-resolution rotary encoders for positioning feedback. An optional linear encoder can also be added to the unit to provide even greater positioning accuracy. The stages utilize crossed-roller slides, precision lead-screws, | ||
| + | The unit can be configured with a number of lead screw options as outlined in the table below. In our standard rotary encoder configuration, | ||
| + | |||
| + | ^Lead Screw Pitch Options ^Rotary Encoder Resolution ^Maximum Speed <wrap lo> | ||
| + | |6.35 mm |88 nm |7 mm/sec| | ||
| + | |1.57 mm |22 nm |1.75 mm/sec| | ||
| + | |0.317 mm |4.5 nm |0.35 mm/sec| | ||
| + | |||
| + | The optional linear encoder provides a resolution of 20 nm and can be used to provide the ultimate feed back device to improve the repeatability and absolute accuracy of the system. Actual resolution & repeatability factors vary somewhat with the load so please contact ASI for these specifications. | ||
| + | |||
| + | The MS-2000 controller accepts industry standard commands, provides automatic backlash correction, and allows RS-232 or USB communication with a host computer. | ||
| + | |||
| + | ===== Adjustable Limit Switches ===== | ||
| + | |||
| + | ==== Newer Dovetail Style ==== | ||
| + | |||
| + | [{{ :: | ||
| + | |||
| + | LS stages are equipped with adjustable magnetic limit switches that can be used to limit the stage travel range. | ||
| + | When the limit magnet is felt by the Hall sensor in the stage body, the limit indicator will show on the controller display and motion will stop in the direction into the limit. Use a 1.5mm allen (or hex) wrench to adjust the limit. | ||
| + | |||
| + | ==== Old Style ==== | ||
| + | |||
| + | [{{ :: | ||
| + | |||
| + | On older linear stages, the limit magnets were held in a series of small holes, kept in place with a 0.05 in set-screw. Limits were adjusted my unscrewing the set-screw, extracting the magnet and placing it in a different hole. When doing this, keep the magnet stuck to the tip of the set screw so it doesn' | ||
| + | |||
| + | |||
| + | |||
| + | |||
| + | |||
| + | |||
| + | ===== Installation Instructions ===== | ||
| + | |||
| + | <WRAP center round important 60%> | ||
| + | WARNING: | ||
| + | When installing the LS stages you must insure that the proper length 1/4-20 screws are used. If the screws are too long, i.e. exceed the thickness of the slider plate, they can press against the adjoining plate and cause serious damage to the stage. The magnet with the white dot is for the //Upper// travel limit (travel away from the motor end of the stage), the red dot magnet for the //Lower// limit. | ||
| + | |||
| + | </ | ||
| + | |||
| + | ==== Attaching Optional Slide Mounting Plate ==== | ||
| + | |||
| + | The optional Top Plate contains more ¼-20 tapped holes and ¼ Allen head counter-bores than are available on the stage slider. | ||
| + | | ||
| + | {{ ls50_stage_8_.jpg? | ||
| + | {{ ls50_stage_9_.jpg? | ||
| + | {{ ls50_stage_10_.jpg? | ||
| + | {{ ls50_stage_11_.jpg? | ||
| + | |||
| + | The LS-50 has less travel than the LS-100, so to access the counter-bores on the connector end of the stage, it is necessary to remove the connector housing. | ||
| + | |||
| + | ==== Attaching Optional Angle Brackets ==== | ||
| + | |||
| + | Gusseted angle brackets are frequently used to rigidly couple multi-axis components. | ||
| + | |||
| + | {{ ls50_stage_12_.jpg? | ||
| + | |||
| + | {{ ls50_stage_13_.jpg? | ||
| + | |||
| + | ==== Cable Connections ==== | ||
| + | |||
| + | LS stages have a DB15 female connector that attaches to the stage cable. | ||
| + | |||
| + | Dual LS stages that are set up for XY operation utilize a split cable. | ||
| + | |||
| + | Single Axis systems, Z-axis/Zoom systems, and Z-axis/ | ||
| + | |||
| + | {{ ls50_stage_14_.jpg? | ||
| + | |||
| + | {{ ls50_stage_1_.jpg? | ||
| + | |||
| + | |||
| + | |||
| + | | ||
| + | {{tag> | ||