| + | ====== Olympus BH-2 Microscope Motor Drive Installation ====== | ||
| + | ===== Motor Drive Installation ===== | ||
| + | |||
| + | |||
| + | This procedure steps you through the installation and alignment of the ASI motor drive onto the Olympus BH-2 microscope. The following tools are required for this procedure: | ||
| + | |||
| + | Small knife or screwdriver \\ | ||
| + | 2.5 mm Allen wrench ( Provided ) \\ | ||
| + | 3.0 mm Allen wrench ( Provided ) \\ | ||
| + | 1.5 mm Allen wrench ( Provided ) \\ | ||
| + | 3/32 mm Allen wrench ( Provided ) \\ | ||
| + | 5/64 mm Allen wrench ( Provided ) \\ | ||
| + | 7/64 mm Allen wrench ( Provided ) \\ | ||
| + | 9/64 mm Allen wrench ( Provided ) \\ | ||
| + | 1/16 mm Allen wrench ( Provided ) \\ | ||
| + | |||
| + | ===== Preparing The Microscope ===== | ||
| + | |||
| + | In this part of the procedure the left fine focus knob will be removed from the microscope: | ||
| + | |||
| + | Prior to removing left fine focus knob, the microscope stage should be fully lowered. To do this, release the lock by rotating fully counter-clockwise the lock ring behind the left coarse focus knob, then rotate the coarse knob to lower the stage all the way down. | ||
| + | |||
| + | <WRAP center round info 60%> | ||
| + | Note: the terms left and right refer to the sides of the microscope as viewed from the front. | ||
| + | </ | ||
| + | |||
| + | {{ bh2_2_.jpg? | ||
| + | |||
| + | 1) Carefully slip the knife point into the inside edge of the left fine focus knob end and pry out the 20mm black disk that covers the fine focus knob attachment screw. | ||
| + | |||
| + | {{ bh2_3_.jpg? | ||
| + | |||
| + | 2) Remove the left fine focus knob by first removing the attachment screw using the 2.5mm hex wrench then pulling the knob off of the fine focus shaft. It will be necessary to hold the right fine focus knob while removing the screw to prevent the shaft from turning. | ||
| + | |||
| + | {{ bh2_washer.jpg? | ||
| + | |||
| + | 3) After the knob has been removed insure that the wave washer is still located on the brass clutch assembly on the focus shaft. | ||
| + | |||
| + | ===== Installing the Drive and Back plate assembly ===== | ||
| + | |||
| + | |||
| + | Locate the motor drive, the back plate and the brass clutch plate. | ||
| + | |||
| + | [{{ bh2_5_.jpg? | ||
| + | |||
| + | [{{ bh2_6_.jpg? | ||
| + | | ||
| + | In this step we will install the ASI backplate and motor drive assembly as a unit. The ASI backplate will be installed and clamped to the back of the microscope as shown only after the motor drive is clamped onto the fine focus shaft of the microscope. Before installing the backplate and drive assembly loosen but do not remove the horizontal and vertical adjustment screws. When these screws are loosened the motor drive should move freely about in the horizontal and vertical adjustment axis. | ||
| + | |||
| + | Alternatively the drive and the back plate can be installed separately. | ||
| + | |||
| + | |||
| + | [{{ bh2_7_.jpg? | ||
| + | |||
| + | [{{ bh2_8_.jpg? | ||
| + | |||
| + | ===== Installing the Drive & Brass Clutch Plate ===== | ||
| + | |||
| + | Use the 7/64 Allen wrench to loosen the clamp that is located on the ASI drive shaft. | ||
| + | | ||
| + | Press the drive onto the microscope until the side of the back plate assembly slides against the side of the microscope, or if installing the drive and back plate separately: push the drive on until the tab on the adjustment bar seats into the grove on the back plate. | ||
| + | ** | ||
| + | |||
| + | [{{ bh2_9_.jpg? | ||
| + | |||
| + | [{{ bh2_10_.jpg? | ||
| + | |||
| + | [{{ bh2_11_.jpg? | ||
| + | | ||
| + | [{{ bh2_12_.jpg? | ||
| + | |||
| + | ===== Aligning the drive ===== | ||
| + | |||
| + | After installing & securing the drive and brass clutch plate to the focus shaft of the microscope the back plate can be positioned and secured. | ||
| + | |||
| + | [{{ bh2_13_.jpg? | ||
| + | | ||
| + | {{ bh2_14_.jpg? | ||
| + | Use the 5/64” Allen Wrench to tighten the Base plate clamp at a point where the vertical adjustment screws are midway in the slots | ||
| + | |||
| + | After securing the back plate check for proper drive alignment be rotating the right fine focus knob. The drive is usually self aligning and when properly aligned no drag is felt, other then that caused by the gears, while rotating the fine focus knob. When the dive is in a position of proper alignment use the 5/32 Allen wrench to tighten the horizontal, and vertical adjustment screws at the position where minimum drag is felt. | ||
| + | |||
| + | Recheck for minimum drag on the fine focus shaft by turning the right fine focus knob. If there is any drag felt at any particular point while rotating the right fine focus knob repeat the alignment procedure by loosening the horizontal, and vertical adjustment screws and repositioning the drive. | ||
| + | |||
| + | |||
| + | ===== Step 6 - Installing the cover & left fine focus knob ===== | ||
| + | |||
| + | {{ bh2_1_.jpg? | ||
| + | |||
| + | Remove the two screws from the edges of the drive assembly. Locate the ASI cover plate. Place it in position over the motor drive assembly and secure it with the screws just removed. Place the left fine focus knob over the protruding end of the drive shaft and secure it in place with the original screw using the 2.5 mm allen wrench. Install the 20mm black disk that covers the fine focus knob attachment screw. | ||
| + | |||
| + | ===== Connecting the motor drive to the controller and computer ===== | ||
| + | |||
| + | |||
| + | Please refer to the wiring drawings and diagrams given in the manual | ||
| + | |||
| + | **Tools Needed:** Small slotted screw driver. | ||
| + | |||
| + | 1) Locate the drive cable. | ||
| + | |||
| + | 2) Locate the computer interconnect cable. | ||
| + | |||
| + | 3) Plug in the power cord / power supply | ||
| + | |||
| + | This completes the steps necessary to interconnect the ASI motor drive, focus controller and computer. | ||
| + | |||
| + | {{tag> | ||