| Line 2: | Line 2: | ||
| <WRAP center round alert 60%> | <WRAP center round alert 60%> | ||
| - | Direct exposure to laser light can lead to an eye injury. During alignment the laser beam may be unpredictable. | + | Direct exposure to laser light can lead to an eye injury. During alignment, the laser beam may be unpredictable. |
| </ | </ | ||
| When aligning a [[tirf_overview|TIRF]] (Total Internal Reflection Fluorescence microscopy) system, one must be considerate of the tilt, and focus of the excitation light. The goal is to have the excitation beam be slightly off-center but perpendicular to the back focal plane of the objective. The TIRF system is aligned if the following are true. | When aligning a [[tirf_overview|TIRF]] (Total Internal Reflection Fluorescence microscopy) system, one must be considerate of the tilt, and focus of the excitation light. The goal is to have the excitation beam be slightly off-center but perpendicular to the back focal plane of the objective. The TIRF system is aligned if the following are true. | ||
| - | - The excitation beam is aligned through | + | - The excitation beam is aligned through |
| - The excitation beam is straight as it passes through the iris and TIRF lens. | - The excitation beam is straight as it passes through the iris and TIRF lens. | ||
| - The excitation beam is straight before the beam splitter, but with a radial offset from center. | - The excitation beam is straight before the beam splitter, but with a radial offset from center. | ||
| - The excitation beam is centered on the non-adjustable axis in the objective mount. | - The excitation beam is centered on the non-adjustable axis in the objective mount. | ||
| - | - The excitation beam is focused to the back focal plane of the objective. | + | - The excitation beam is focused to the back focal plane of the objective |
| - | **1) Fiber alignment: | + | **1) Fiber alignment: |
| - | **2) Condenser alignment: | + | {{ : |
| + | |||
| + | **2) Condenser alignment: | ||
| **3) TIRF assembly alignment: | **3) TIRF assembly alignment: | ||
| - | |||
| - | **4) Beam splitter alignment: | ||
| - | |||
| - | **5) TIRF lens adjustment: | ||
| - | |||
| - | |||
| - | This manual covers installing and aligning ASI's manual and motorized TIRF assembly on a RAMM. If you are looking for instruction on how to align the TIRF, then jump to [[tirf_alignment# | ||
| - | |||
| - | <WRAP center round alert 60%> | ||
| - | Direct exposure to laser can lead to an eye injury. Please follow all safety precautions. | ||
| - | </ | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | The assembly consists of a | ||
| - | - Fiber launch | ||
| - | - Actuator to position the fiber tip | ||
| - | - Beam expander | ||
| - | - Lens cage assembly - allows users to focus the beam onto the Back Focal Plane (BFP) of the objective | ||
| - | |||
| - | ====== Assembling the TIRF ====== | ||
| - | |||
| - | For safer shipping, ASI might ship the TIRF disassembled. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Install the focusing lens by lining up the four holes in the lens' ring mount with the four stainless steel rods and slide it in with the collar of the ring facing out / towards the RAMM. The curved side of the lens now points towards the objective. The ring has four 1.5 mm set screws; you might have to loosen them so they can slide on the rods. Leave the set screws loose for now. | ||
| - | |||
| - | The picture above shows a TIRF with 200 mm lens. Depending on the configuration of your particular RAMM or microscope, ASI might ship you a different focal length lens. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | If the actuator is shipped separately, attach it by removing the plastic cover on the fiber launch by unscrewing the four Phillips head screws. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Insert the actuator all the way. Orient the micrometer such that the graduation markings are easy to read. Then lock it in place by tightening the set screw with a 1.5 mm hex wrench. Replace the cover. | ||
| - | |||
| - | |||
| - | ====== Limiting the travel range of motorized actuator ====== | ||
| - | |||
| - | [{{ : | ||
| - | |||
| - | A note of caution for TIRF with the motorized actuator, the [[http:// | ||
| - | |||
| - | [{{ : | ||
| - | |||
| - | We suggest limiting the MA12's travel range by setting the soft upper limit at a position that doesn' | ||
| - | |||
| - | ====== Attaching the TIRF assembly to RAMM ====== | ||
| - | |||
| - | TIRF is commonly connected to a Cube 2, Cube 3, or a motorized cube. The cubes have a ring with four holes for the cage rods already installed. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | First, open the C clamp support by unscrewing the two screws shown in the above image with a 2.5 mm hex wrench. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Next, slide the assembly thru it, and line up the post with the corresponding holes and push it in. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Then secure the stainless steel posts to the ring by tightening the four set screws with a 1.5 mm hex wrench. Location of the four set screws shown in the above pictures. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Then put the support clamp back together so that the TIRF assembly is well supported. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Lastly, the actuator should be oriented on the left side and angled such that its parallel to the table or optical bench. Loosen the fiber launch set screw with a 2 mm hex wrench and orient it. This is not an absolute rule; it is safest to have the collimated beam exiting the objective tilt away from users. | ||
| - | |||
| - | ====== Aligning the TIRF ====== | ||
| - | |||
| - | <WRAP center round alert 60%> | ||
| - | Direct exposure to laser can lead to an eye injury. During alignment the laser beam may be unpredictable. | ||
| - | </ | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Connect the laser to the fiber launch. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | To verify the laser is correctly coupled to the fiber launch, unscrew the three black thumb screws and remove the CUBE's cover. Careful to not look at the laser directly, but safely check for its presence. If you do not see the laser exit the cube, there might be a problem with your fiber or the fiber launch on the light source. | ||
| - | |||
| - | If you haven' | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Position the fiber launch at the center of the tube. If you have a motorized actuator, assign the actuator to a manual input device (joystick or knob) with the [[commands: | ||
| - | |||
| - | Adjust the tilt of the dichroic by rotating the stainless steel thumb screws so that the beam exits the objective vertically and hits the ceiling right above the objective. | ||
| - | |||
| - | Be careful and do not look at the objective directly; as you make adjustments the beam may exit the objective at odd angles. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Initially, the beam will probably be dim and broad. Use the caged lens to focus the beam onto the BFP of the objective. Adjust the position of the tube lens in the cage assembly until you see a small bright dot. When satisfied, lock the position of the lens by tightening one or two of the four setscrews in the lens ring with a 1.5 mm hex wrench. | ||
| - | |||
| - | The steps so far covered basic alignment; the steps below get into TIRF mode. For this, we need a sample on the stage with the objective at focus. | ||
| - | |||
| - | [{{ :: | ||
| - | |||
| - | Initially, the beam exits the objective and goes through the sample vertically. | ||
| - | |||
| <WRAP center round info 60%> | <WRAP center round info 60%> | ||
| - | Tip for people with the motorized actuator option, you may want to save this position of the TIRF actuator as it lets you do widefield imaging. Save this position | + | While rotating |
| </ | </ | ||
| + | {{ : | ||
| - | [{{ ::tirf21.jpg?direct&400 |Steer the beam by rotating | + | **4) Beam splitter alignment: |
| - | + | ||
| - | [{{ ::tirf28.jpg? | + | |
| - | + | ||
| - | As you continue to move the fiber launch from its central position by rotating | + | |
| - | + | ||
| - | [{{ ::tirf29.jpg?direct&400 |Beam shooting out horizontally}}] | + | |
| - | + | ||
| - | Continue moving | + | |
| - | Motorized | + | **5) TIRF lens adjustment:** Mount the objective on the scope. Place a light blocking object between your eyes and the tip of the objective, such as an open book, to act as a beam-stop for safety. Turn the laser on and open the iris. Adjust the micrometer until the beam is shining on the ceiling. If the beam makes an Airy disc pattern you are done. If not, [[optical_and_mechanical_alignment_on_ramm# |
| Your TIRF assembly alignment is now complete. For questions and clarifications email ASI at '' | Your TIRF assembly alignment is now complete. For questions and clarifications email ASI at '' | ||
| + | For the old assembly and alignment procedures, see [[tirf_alignment_old|the old TIRF documentation]]. | ||